Evaluation of the use of PEEK material in a knee joint endoprosthesis insert by FEM analysis

Wiktoria M. Wojnarowska, Slawomir Miechowicz and Tomasz Kudasik

Abstract

This study contains a comparative analysis of two polymers: PEEK and UHMWPE. A stress distribution pattern in a knee joint endoprosthesis insert was determined for both materials. A numerical model of the insert generated by means of the SOLIDWORKS software was used in calculations. For the purpose of the calculations it was assumed that the insert is loaded consistently and symmetrically, and the value of the load applied was determined for a person of a total weight of 120 kg.

Key Words

endoprosthesis; finite elements methods; polyetheretherketone; implant devices

Address

Wiktoria M. Wojnarowska: Doctoral School of Engineering and Technical Sciences at the Rzeszow University of Technology, Poland Slawomir Miechowicz: Department of Mechanical Engineering, Faculty of Mechanical Engineering and Aeronautics, Rzeszow University of Technology, Poland Tomasz Kudasik: Department of Mechanical Engineering, Faculty of Mechanical Engineering and Aeronautics, Rzeszow University of Technology, Poland